Download Print this page

Frequency Band Ghz - HP 8565A Manual Change

Hide thumbs Also See for 8565A:

Advertisement

Model8565A
General Information
Table 1-1. Model8565A Specifications (2 of 3)
Average noise level:
Sensitivity (minimum discernible signal) is given by the signal
level which is equal to the average noise level, causing approx-
imately a 3 dB peak above the noise. Maximum average noise
level with 1 kHz Resolution Bandwidth (0 dB attenuation and
0.003 (3 Hz) video filter) is given in the table below:
Frequency
First IF
Harmonic
Avg. Noise Level
Band (GHz)
inMHz
Mode
(dBm)
.01-1.8
2050
1-
-112
1.7-4.1
321.4
1-
-109
3.8-8.5
321.4
2-
-103
5.8-12.9
321.4
3-
-94
8.5-18
321.4
4+
-87
10.5-22
321.4
5+
-75
Reference Level
Reference Level range:
+
70 dBm1 to -102 dBm .in 10 dB
steps and continuous 0 to -12 dB calibrated vernier.
Reference Level accuracy:
With Sweep Time/Division control in Auto setting, the opti-
mum sweep rate is selected automatically for any combin-
ation of Frequency Span/Div, Resolution Bandwidth and
Video Filter settings. Thus, the Auto Sweep setting insures
a calibrated amplitude display within the following limits:
Calibrator output
-10dBm ±0.3 dB
100 MHz ±10kHz
Reference Level variation (Input Attenuator at 0 dB)
10 dB Steps:
<
±0.5dB(Oto -70dBm)
<
± 1.0 dB (0 to -90 dBm)
Vernier (0 to -12 dB) continuous: Maximum error
<±o.5 dB, when read from Reference Level Fine control.
Input Attenuator (at preselector input, 70 dB range in 10
dB steps)
Step size variation (for steps from 0 to 60 dB):
<
±1.0dB
.01 to 18GHz
Maximum cumulative Error (from 0 to 60 dB):
<
± 2.4 dB
.01 to 18 GHz
Frequency Response (with 0 or 10 dB of Input Attenuation)
Frequency response includes input attenuator, preselector
and mixer frequency response plus mixing mode gain var·
iation (band to band).
Frequency Band (GHz)
0.01 to 1.8
1.7 to 4.1
3.8 to 8.5
5.8 to 12.9
8.5 to 18
10.5 to 22
Frequency Response
(±dB MAX.)
1.2
1.7
2.5
2.5
3.0
4.5
1 Input level not to exceed+ 30 dBm damage level.
'2
100 kHz BW limited to 9007o relative humidity.
CHANGE 3 6 DECEMBER 1977
Switching between bandwidths: 3 MHz to 300 kHz, ±0.5
dB; 3 MHz to 1 kHz, ± 1.0 dB2.
Calibrated display range
Log -expanded from reference level down:
70 dB with 10 dB/Div scale factor
40 dB with 5 dB/Div scale factor
16 dB with 2 dB/Div scale factor
8 dB with 1 dB/Div scale factor
Linear. Full scale from 56 p.V ( -102 dBm across 50 ohms
to 707 volts (+70 dBm)1 in 10 dB steps and continuous 0 to
-12 dB vernier. Full scale signals in linear translate to ap-
proximately full scale signals in the log modes.
Display accuracy
Log:
<
±0.1 dB/dB but not more than ±1.5 dB over full
70 dB display range.
Linear:
<
± 0.1 division over full 8 division deflection
Residual responses (no signal present at input):
With 0 dB input attenuation on fundamental mixing (0.01 to
4.1 GHz)
<
-90 dBm.
Signal Identifier: Provided on all Frequency Bands in
1 MHz/Div frequency span for signal identification (reads
incorrectly for 100 MHz Cal Output signal)
SWEEP SPECIFICATIONS
SWEEP TIME
Auto: Sweep time is automatically controlled by Frequency
Span/Div, Resolution Bandwidth and Video Filter controls to
maintain an absolute amplitude calibrated display.
Calibrated Sweep times: 21 internal sweep times from 2
p.sec/Div to 10 sec/Div in 1, 2, 5, 10 sequence. Sweep time
accuracy ±10% except for 2, 5, and 10 sec/Div which are
± 20%. Swept frequency modes use sweep times 2 msec/Div
through 10 sec/Div. When operated as a fixed tuned receiver
(Zero Span) the full range of sweep times(2 p.sec to 10 sec/
Div) may be used to display modulation waveforms.
GENERAL SPECIFICATIONS
TEMPERATURE RANGE:
Operating 0°C to 55°C, storage -40° to
+
75°C.
HUMIDITY RANGE (Operating):
<
95% R.H. 0°C to 40°C.
EM I:
Conducted and radiated interference is within the require-
ments of methods CE03 and RE02 of MIL STD 461A, VDE 0871
and CISPR pub'n 1, 2, and 4.
POWER REQUIREMENTS
48-66Hz, 100, 120, 200 or 240 volts ( -10% to +5%) 220 VA
maximum . Fan cooled.
1-3

Advertisement

loading