CAUTION
CAUTION
Calibrating for Increased Measurement Accuracy
Calibration Choices
To Perform a Normalization Calibration
Normalization is the simplest type of calibration. The analyzer stores
data into memory and divides subsequent measurements by the stored
data to remove frequency response errors.
Follow these general steps when performing a normalization calibration:
1. Set up the analyzer for your measurement:
• Select the type of measurement.
• Enter operating parameters other than the default.
2. Connect your equipment as you would for an actual measurement,
but omit the DUT.
3. Press
DISPLAY
4. Notice that the top of the display on the analyzer now shows the type
of measurement followed by "/M" to indicate that the displayed data
is actually the measurement divided by memory.
The normalization cal will be invalidated if any frequency settings are
changed after calibration. There is no correction interpolation with a
normalization cal.
If you decide to perform any other type of calibration, be sure to turn off
normalization first by pressing
Normalize ON off
normalization is turned off.
6-14
Normalize
or
CAL
DISPLAY
. Note that the "/M" indicator disappears when
Normalize on OFF
Data
or
CAL
ES User's Guide
.