Softkey in the user TTL config menu. Pressing this softkey configures
the rear panel USER TTL IN/OUT connector for general purpose I/O
use with IBASIC and SPCI commands. This connector can also be
configured for softkey sequencing using an external switch, and as a
sweep out connector.
See "User-Defined TTL Input/Output" in the Automating
Measurements User's Guide Supplement, and
Connectors—Rear Panel BNC Connectors" on page 7-4
information.
Access Keys:
SYSTEM OPTIONS
User TTL Config
Softkey in the
CAL
in S
, or S
(reflection) measurement mode. Use this key to recall a
11
22
one-port calibration that was performed at the factory or during
servicing. Corrects for directivity, source match, and frequency
response errors.
See
Chapter 6
for more information on calibrating the analyzer.
Softkey in the color options menu. Sets the external VGA monitor
colors to the Default 2 setting.
See
"Using an External VGA Monitor" on page 4-87
information.
Access Keys:
DISPLAY
Softkey in the
CAL
S
, S
, or S
(reflection or transmission) measurement mode. Use
21
12
22
this key to recall a two-port calibration that was performed at the
factory or during servicing. Corrects for directivity, source match, load
match and frequency response errors. Also corrects for crosstalk errors
when the
Isolation on OFF
See
Chapter 6
for more information on calibrating the analyzer.
ES User's Guide
System Config
menu and the one-port reflection cal menu when
More Display Color Options
menu and the two-port cal menu when in S
key is toggled to ON.
Hardkey/Softkey Reference
"Analyzer
for more
for more
11