To Turn Off Alternate Sweep

Alternate sweep is turned off when the analyzer is preset, but is
automatically activated with some dual channel measurements. The
alternate sweep feature sweeps and measures one channel at a time. By
disengaging this feature, you increase the sweep speed by a factor of two.
1. Press
SWEEP
2. If ON is in capital letters, toggle the softkey to disengage the
alternate sweep feature (the softkey should appear
Alt Sweep on OFF

To Turn Off Markers and Marker Tracking

When markers are on, time is required to update the marker readouts.
Turning off markers can reduce sweep cycle time by up to 30 ms. When
the marker softkey menu is active, it too must be updated, adding up to
10 ms to the sweep cycle time.
Press
All Off
MARKER
When marker tracking is on, the analyzer performs a search after each
sweep. Simple searches like
sweep. Lengthy calculations like
each sweep.
1. Press
MARKER
Tracking on OFF
2. If ON is in capital letters, toggle the softkey to the OFF position.
ES User's Guide
and look at the
Alt Sweep on OFF
).
.
Mkr -> Max
Bandwidth
Marker Search
softkey.
Optimizing Measurements
Increasing Sweep Speed
softkey.
can take 10 to 20 ms per
can add over 100 ms to
and look at the
5-7