HP 8712ES User Manual page 169

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

Delta Amplitude
Delta Frequency
This marker limit test allows you to set marker 1 as an amplitude
reference against which marker 2 is limit tested.
1. This limit test requires that you first use marker 1 to determine the
reference amplitude: Press
panel knob or the
place on the measurement trace. The amplitude of marker 1 at this
point becomes the delta reference for this marker limit test.
2. Press
DISPLAY
3. Use the front panel knob or the
Ampl in the marker limit test table. Turn this limit function on by
pressing the
Mkr Limit on OFF
on/off column of the table changes to "on."
4. Press
Edit Limit (Min/Max)
maximum limit using the front panel keypad and terminating the
entry with the
ENTER
5. Press
Prior Menu Edit Limit (Min/Max)
enter the minimum limit using the front panel keypad and
terminating the entry with the
This marker limit test allows you to set marker 1 as a frequency
reference against which marker 2 is limit tested.
1. This limit test requires that you first use marker 1 to determine the
reference frequency: Press
panel knob or the
place on the measurement trace. The frequency of marker 1 at this
point becomes the delta reference for this marker limit test.
2. Press
DISPLAY
3. Use the front panel knob or the
Freq in the marker limit test table. Turn this limit function on by
pressing the
Mkr Limit on OFF
on/off column of the table changes to "on."
4. Press
Edit Limit (Min/Max)
maximum limit using the front panel keypad and terminating the
entry with the
ENTER
5. Press
Prior Menu Edit Limit (Min/Max)
enter the minimum limit using the front panel keypad and
terminating the entry with the
ES User's Guide
MARKER
keys to place marker 1 at the desired
Limit Menu
Mkr Limits
key. Note that the entry in the
Max Limit
key.
ENTER
MARKER
keys to place marker 1 at the desired
Limit Menu
Mkr Limits
key. Note that the entry in the
Max Limit
key.
ENTER
Using Instrument Functions
Using Limit Testing
1:
, and then use the front
.
keys to select Delta
, and enter the
Min Limit
, and
key.
1:
, and then use the front
.
keys to select Delta
, and enter the
Min Limit
, and
key.
4-35

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hp 8714es

Table of Contents