Attenuation And Amplification In A Measurement Setup - HP 8712ES User Manual

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

CAUTION
NOTE
Attenuation and Amplification in a
Measurement Setup
The measurement setup that you use may require attenuation or
amplification. The following sections describe when to use them.
When to Use Attenuation
• For accurate measurements, use external attenuation to limit the
power at the input port to +10 dBm (for narrowband-detection
measurements) or +16 dBm (for broadband-detection measurements).
Always use attenuation on the input port if your test device's output
power exceeds the receiver damage limit of +26 dBm or ±30 Vdc.
• For information on how to reduce mismatch errors, see
Mismatch Errors" on page
When to Use Amplification
• For accurate measurements, amplification may be needed on the
analyzer's output port. Use amplification when your test device
requires input power that exceeds the analyzer's maximum specified
output power.
If you use an amplifier between the analyzer's output port and your DUT,
you won't be able to measure the input match (S
The maximum specified output power is dependent upon the option
configuration of your analyzer as well as the frequency range of your
test setup. It ranges from +4 to +13 dBm. See
"Specifications,"
to determine the maximum specified output power of
your analyzer.
ES User's Guide
Measuring Devices with Your Network Analyzer
5-15.
11
Chapter 9,
Making Measurements
"Reducing
) of your DUT.