Making Measurements
Measuring S-Parameters using a Two-Port Calibration
20. Start the calibration process by pressing
21. When the instrument prompts you to connect a through cable, insert
Figure 3-14
Through-Cable Connection
22. Remove the adapter between the test cables connected to port 1 and
3-38
If you want to enter a cal kit description in the display's table,
press
Add Kit Description
analyzer's DIN KEYBOARD connector to enter the description.
(If you don't have a keyboard, you can use the analyzer's
keys or the front panel knob and the
key to select and enter characters from the top of the analyzer's
Enter
display. Press
Reflection Cal menu is displayed. Next, press
User 2-Port
. If you want to correct for crosstalk errors, toggle ON
the
Isolation on OFF
refer to the definition of a user two-port calibration in
the type-N-to-3.5 mm female-to-female adapter between the test
cables connected to port 1 and port 2 as shown in
Measure Standard
port 2.
. Use a keyboard connected to the
when done.)
Prior Menu
key (for more information about isolation,
.
Select Char
until the
User 2-Port
Chapter
6).
Figure
3-14. Press
ES User's Guide