An optimum calibration is critical for achieving best measurement
accuracy. Refer to
Chapter 6, "Calibrating for Increased Measurement
Accuracy,"
for detailed information about all of the various aspects of
calibration.
Phase-derived delay cannot be used to measure frequency-translating
devices.

Enter the Measurement Parameters

1. Press the following keys on the analyzer:
PRESET
(
User Preset
MEAS 1
S21 Fwd Trans
FORMAT
Delay
Center
FREQ
Span
200
2. Choose an aperture. When choosing an aperture, there is a trade-off
between minimum apertures (giving more resolution but noisier
responses) and maximum apertures (giving less resolution but
smoother responses). For this example, choose an aperture of 4% by
pressing:
AVG
Delay Aperture
Aperture (%)
Enter
4
3-70
or
Factory Preset
MHz
175
MHz
)
ES User's Guide