Enter The Measurement Parameters; Perform A User One-Port Calibration - HP 8712ES User Manual

Rf network analyzers
Hide thumbs Also See for HP 8712ES:
Table of Contents

Advertisement

NOTE

Enter the Measurement Parameters

Press the following keys on the analyzer:
PRESET
(
User Preset
or
MEAS 1
S11 Refl Port1
This example measurement uses the default instrument parameters for
an S
measurement. If your particular reflection measurement requires
11
different parameters (such as frequency range, source power level,
number of data points, and sweep time), enter them now.

Perform a User One-Port Calibration

This example describes how to perform a one-port reflection calibration.
A one-port reflection calibration uses known standards to correct for
directivity, source match, and frequency response errors when using
narrowband detection. When you perform a one-port reflection
calibration, the analyzer performs correction at each data point across
the selected frequency band. The default number of data points per
sweep is 201, but you can select any number between 3 and 1601.
Interpolation recalculates the error correction array for reduced
frequency spans. If the frequency span is increased the calibration is
invalidated, and the default one-port calibration is automatically
restored.
ES User's Guide
Measuring S11 Reflection Port 1 using a One-Port Calibration
Factory Preset
)
Making Measurements
3-49