Configuring the measured values line
Select one of 6 possible combinations of measured values:
∆∆
Press »Config.
« key repeatedly until
»Configuration 1/4« appears.
Select »Measured values« line = turn the rotary knob,
activate = press the rotary knob.
To set the measured values combination =
turn the rotary knob,
confirm = press the rotary knob.
The following combinations of measured values can
be selected:
1.
VTe
2.
FiO
2
3.
Ppeak
4.
Pmean
5.
Pmean
6.
Ppeak
Manual calibration for O
— once per month,
— if the following alarms occur:
!!! FiO
high
2
or
!!! FiO
low
2
or
!!! O
measurement inop.
2
Manual calibration can be carried out in the respective
ventilation mode:
∆∆
Press »Config.
« key repeatedly until
»Configuration 2/4« appears.
Select »O
calib.« line = turn rotary knob,
2
to release for setting = press rotary knob.
To start O
calibration = turn rotary knob and select
2
»Start«,
to confirm = press rotary knob.
Message in the information line on screen:
Disconnect patient
Disconnect patient within the next 30 seconds.
If necessary continue ventilating the patient using a
separate ventilating appliance.
ftot
MV
VTe
MV
VTe
MV
FiO2
MV
VTe
MV
Pmean
V
Te
sensor 2
2
Configuration
49