4 Operation | 4.12 Using the KL 703 LED in ENDO mode (optional accessory)
S1
S2
S3
Parameters
Transmission factor
Parameter memory
Torque mode
Direction of motor rotation
Speed
Torque
Incorrectly set parameters.
Injury or property damage from incorrect input values.
▶ Check all input values before use.

4.12.3 Set torque mode

Three different torque modes are available:
▪ Autoreverse
▪ Torque Control only
▪ Autorev / Forward
▶ Press the "Up" or "Down" key to select the desired torque mode.
ð
The torque mode is shown on the display and is effective immediately.
Torque mode Torque Control only
▶ Press the foot pedal.
ð
The motor starts by rotating clockwise (if not selected otherwise).
The torque is limited to the set threshold. The speed reduces until it stops de-
pending on the load.
The direction of rotation is always to the right.
A tone sounds when the set torque threshold is reached.
S4
Values
1:1, 3:1
P1,P2,P3,P4,P5,P6
TQ-Ctrl, Autoreverse, AutoRev/Forw.
R, L
100 to 6,000 rpm
Transmission ratio 1:1: 0.15 to 2.50
Ncm
Transmission ratio 3:1: 0.4 to 8.0 Ncm
CAUTION
95 / 134
Instructions for use ESTETICA E30