Instructions for use ESTETICA E30
4 Operation | 4.3 Adjusting the dental chair
▶ Backrest down: Move the cross switch on the foot control in direction ④.

4.3.4 Automatic positioning of dental chair

Danger of injury from overload or dynamic load.
The patient chair might collapse.
▶ Do not subject the patient chair to a load exceeding its limit (180 kg).
▶ Do not subject the patient chair to dynamic loads.
Danger of crushing during automatic chair movement.
The patient or treatment personnel can be clamped.
▶ Monitor the patient and treatment personnel when changing the chair posi-
tion.
The chair position can be adjusted continuously.
Automatic positions can be saved, and the saved positions can be recalled by
the push of a button.
Recalling automatic positions with the dentist unit
The following keys can be used to recall saved chair positions.
Key
▶ Briefly press the desired button.
ð
Chair automatically moves to the stored position.
ð
Upon arrival at the stored position, the display diode on the button is turned
on.
Saving automatic positions with the dentist unit
Recommended assignment of buttons:
"SP" button: rinsing position
"AP 0" button: entry and exit position
CAUTION
CAUTION
Operation
Move to the rinsing position.
The last position before actuating the SP is assumed.
Move to automatic position 0.
Move to automatic position 1.
Move to automatic position 2.
Move to the collapsed position.
48 / 134