Pos.
Safety switch-off ac-
no.
tuated
Bracket on the foot
control
Assistant element
Backrest
Kickplate
Seat
For right/left conver-
sion with bench re-
moved
The safety shutoff occurs if a movement angle has been exceeded or if a part of
the treatment centre collides with an object.
If a person or object actuates a safety shutoff, the chair immediately stops
moving.
The fact that the safety shutoff has been activated is displayed by the corres-
ponding display flashing on the dentist or assistant element.
Note
The chair's position cannot be changed with the key wheels when a safety
shutoff is activated.
▶ To deactivate an activated safety shutoff, remove the triggers from the mo-
tion range of the chair.
Changing the chair's position when the safety circuit is on.
Personal injury.
Damage to the device.
▶ Changing the chair position while a safety shutoff is active, do not move the
chair against the active safety circuit.
Pinching from the treatment chair.
The safety shutoff of the treatment chair is activated by lifting the respective
component. Depending on the patient's body weight and the leverage, more
force can be exerted on the object to be triggered than is necessary to trigger
the switching function.
▶ The treatment personnel must move outside of the chair's swivelling range
whenever the chair moves.
To allow the chair to move freely, it can also be moved when the safety circuit
is on. Use this function for repair purposes only.
▶ Press and hold down both the "SP" and "LP" keys on the dentist element.
or
4 Operation | 4.3 Adjusting the dental chair
LED on assistant
element
CAUTION
CAUTION
53 / 134
Instructions for use ESTETICA E30
LED on dentist ele-
ment