Electrical power.
Electrical shock.
▶ Set up the external PC outside of the patient environment keeping a min-
imal distance of 1.5 m.
▶ Connect the PC and equipment connected to the PC in accordance with IEC
60601-1 / 60950.
Electrical power.
Electrical shock from incorrectly connecting a non-medical system to the USB
interfaces of the device.
▶ Connect any IT device to the medical system in accordance with IEC
60601-1.
▶ Use USB devices with no additional power supply (USB-powered) only.
▶ Applied parts connected to the USB interface of the dentist element must
comply with the requisite insulation.
▶ USB-powered devices failing to meet the requisite insulation for applied
parts must be placed appropriately such that direct contact of the USB
device and the patient is excluded.
▶ It is not permissible to touch USB-powered devices failing to meet the re-
quisite insulation for applied parts and the patient at the same time.
Health damage due to germ formation.
Infection hazard.
▶ Before start-up and after downtimes (weekends, holidays, vacations, etc.),
rinse or purge the air and water lines with air.
▶ Actuate the tumbler filler several times.
▶ Carry out a manual intensive germ reduction.
Long stay in the patient chair.
Decubitus formation.
▶ Take precautions against the formation of decubitus in long treatments.
Risk of injury when the dental chair or headrest is moved.
Hair of the patient or practice personnel may get caught when the headrest of
the dental chair is moved.
▶ Mind the hair of the patient or practice personnel when moving the dental
chair or the headrest.
Damage to the instrument hoses from stickers.
Instrument hoses can burst.
▶ Do not affix stickers or adhesive tape.
Risk of injury when the dentist or assistant element is moved.
The patient or office staff may be injured or bruised.
▶ Monitor the patient and office staff when moving the dentist or assistant
element.
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
NOTICE
CAUTION
17 / 134
Instructions for use ESTETICA E30
2 Safety | 2.3 Safety instructions

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents