Instructions for use ESTETICA E30
2 Safety | 2.3 Safety instructions

2.3.2 Product-specific

Injury or infection hazard from laid down instruments.
Given the arrangement of the instruments, injury or infections in the hand and
underarm may occur when reaching for the tray holder or operating device. In-
creased risk of infection from ill patients.
▶ Be aware of the arrangement of the instruments when you reach for the
tray holder or operating device.
Health damage due to back-suction on the instruments.
Infection hazard.
It is feasible to use, on standardised interfaces, products from other manufac-
turers, which are not equipped with a protective device preventing back-suc-
tion of treatment water via the instruments into the dental treatment centre.
▶ If products from other manufacturers are used at the standardised inter-
faces, you must make sure that the products are equipped with an appro-
priate protective device!
▶ Do not use products with no protective device.
Sitting down on a dental chair that is in horizontal orientation is asso-
ciated with a risk of injury.
▶ Do not sit on the head or foot end of the patient chair when it is in a hori-
zontal position.
Risk of injury if you support yourself on the swivel arm.
Exposing the swivel arm to an overload may damage the swivel arm and cause
injury to the patient or user.
▶ Never strain the swivel arm, spring arm, assistant element and dentist ele-
ment by using it as a support!
Risk of injury by suspended instruments (S table).
Patients may get injured by sharp instrument tips.
▶ When you move the dentist's unit, make sure that nobody is injured.
▶ Alert patients and care providers to the risk of injury.
Risk of injury during cleaning of the treatment unit.
Lack of instructions to the cleaning staff and lack of preparation of the treat-
ment unit can lead to the cleaning personnel sustaining injuries.
▶ Only trained professionals and instructed cleaning personnel may be
present in the treatment rooms.
▶ Position the chair for cleaning and turn the device off.
WARNING
WARNING
CAUTION
CAUTION
CAUTION
CAUTION