  • Page 1 Care instructions ESTETICA E30...
  • Page 2 Manufacturer: Distributed by: Kaltenbach & Voigt GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach www.kavo.com Phone +49 (0) 7351 56-0 Fax +49 (0) 7351 56-1488...
  • Page 3: Table Of Contents

    1.1.3 Target group ...................... 4 2 Reprocessing methods according to DIN EN ISO 17664 ........... 5 2.1 Instructions........................5 2.2 General instructions for the reprocessing of KaVo treatment centres ........6 2.2.1 Overview ESTETICA E30 reprocessing..............7 2.2.2 Reprocessing KaVo Screen One / KaVo Screen HD ..........8 2.3 Reprocessing of the surfaces of treatment centre / upholstery ..........
  • Page 4: User Instructions

    1 User instructions 1.1 User guide Requirement Read these instructions prior to first use to avoid misuse and prevent damage. Note Comply with the Instructions for Use of the ESTETICA E30 treatment unit! 1.1.1 Abbreviations Explanation brevi- ation Instructions for use...
  • Page 5: Reprocessing Methods According To Din En Iso 17664

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.1 Instructions 2 Reprocessing methods according to DIN EN ISO 17664 2.1 Instructions The reprocessing consists of the following basic steps: ▪ Cleaning and disinfection ( manually or automatically) ▪...
  • Page 6: General Instructions For The Reprocessing Of Kavo Treatment Centres

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment centres 2.2 General instructions for the reprocessing of KaVo treatment centres Please comply with the general reprocessing instructions at all times. Detailed product-specific reprocessing instructions are provided subsequently.
  • Page 7: Overview Estetica E30 Reprocessing

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.2 General instructions for the reprocessing of KaVo treatment centres Manual drying ▶ Supply dry, clean and uncontaminated compressed air according to ISO 7494-2 only. ▶ Blow off the outside and inside with compressed air until water drops are no longer visible.
  • Page 8: Reprocessing Kavo Screen One / Kavo Screen Hd

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.3 Reprocessing of the surfaces of treatment centre / upholstery 2.2.2 Reprocessing KaVo Screen One / KaVo Screen HD Cleaning and disinfection NOTICE Non-colour-fast textiles. Non-colour-fast textiles can cause permanent discoloration.
  • Page 9 Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.3 Reprocessing of the surfaces of treatment centre / upholstery Cleaning and disinfecting the surfaces of treatment unit / upholstery NOTICE Non-colour-fast textiles. Non-colour-fast textiles can cause permanent discoloration.
  • Page 10: Preparation Of The Dentist Element

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element 2.4 Preparation of the dentist element 2.4.1 Reprocessing of handle and holder pad/anti-slip Note The handle of the TM table cannot be detached.
  • Page 11: Reprocessing Of The Instrument Arm S Table

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element Sterilize the instrument holder NOTICE Damage caused by sterilisation. ▶ Do not sterilise the instrument holder. 2.4.3 Reprocessing of the instrument arm S table Note The swing-arm S-table (Mat.
  • Page 12: Preparation Of The Triple Function Handpiece Kavo One

    Note For the cleaning/preparation of the cannula of the triple-function handpiece KaVo One, the cannula must be disassembled into the cannula tube and the taper sleeve. Cleaning of triple-function handpiece KaVo One Automatic cleaning NOTICE Damage caused by improper cleaning/disinfection.
  • Page 13 ▶ Carefully remove contamination from the media exit ports using the short nozzle cleaning needle. Disinfection of triple-function handpiece KaVo One Manual disinfection See also: 2 2.2 General instructions for the reprocessing of KaVo treatment centres, Page 6 External disinfection ▶ Wipe-disinfect the surfaces with a soft cloth. Automatic disinfection NOTICE Damage caused by improper cleaning/disinfection.
  • Page 14: Preparation Of Triple Function Handpiece

    Note Wet the O-rings of the gripping sleeves after sterilisation using silicone grease (Mat. no. 1.000.6403) or KaVo Rota Spray 2 (Mat. no. 0.411.7510). After replacing the cannula, the air channel must be blown out before the start of treatment to ensure that the air exiting during treatment is dry.
  • Page 15 Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element NOTICE Damage caused by improper cleaning/disinfection. ▶ Do not wash the gripping sleeve and the cannula in the washer disinfector.
  • Page 16 Requirement The multifunctional handpiece has been prepared. ▶ Apply KaVo silicone grease (Mat. no. 10006403) or KaVo Rota Spray 2 (Mat. no. 04117510) to the O-rings of the joint between the gripping sleeve and cannula. Use cotton buds for this purpose.
  • Page 17: Preparation Of The Instrument Hoses

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element ▶ Pull the gripping sleeve off the valve body. Sterilization in a steam sterilizer in compliance with DIN EN 13060 ▶...
  • Page 18: Servicing The Turbine Return Air Filter

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element 2.4.7 Servicing the turbine return air filter Cleaning the turbine return air filter Note The liquid collector must be checked weekly.
  • Page 19 Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element Cleaning Note Disinfect the camera before cleaning. Disinfect after cleaning if necessary. Exterior cleaning by hand NOTICE Damage to the camera due to improper cleaning.
  • Page 20: Reprocessing Kavo Piezoled

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element Approved disinfectants: ▪ CaviCide / CaviWipes (Metrex) ▪ Microcide AF Liquid (Schülke & Mayr) ▪ FD 322 (Dürr Dental) ▪ Incidin Liquid (Ecolab)
  • Page 21 ▶ For programme settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the washer disinfector. ▶ In order to prevent negative effects on the KaVo medical device, make sure that the interior and the exterior of the medical device are dry after comple- tion of the cycle, and then lubricate it immediately with care products from the KaVo care system.
  • Page 22 Manual disinfection - external KaVo recommends the following products based on material compatibility. The microbiological efficacy must be ensured by the disinfectant manufacturer. ▶ Attach disposable syringe (at least 50 ml) to the nozzle of the product.
  • Page 23 ▶ For programme settings as well as cleansers and disinfectants to be used, please refer to the Instructions for Use of the washer disinfector. ▶ In order to prevent negative effects on the KaVo medical device, make sure that the interior and the exterior of the medical device are dry after comple- tion of the cycle, and then lubricate it immediately with care products from the KaVo care system.
  • Page 24 Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.4 Preparation of the dentist element Drying Drying of handpieces Manual drying ▶ Blow off the outside and inside with compressed air until water drops are no longer visible.
  • Page 25 Sterilisation in a steam steriliser (Autoclave) EN 13060/ ISO 17665-1 NOTICE Premature wear and malfunctions from improper servicing and care. Reduced product life. ▶ Before each sterilisation cycle, service the medical device with KaVo care products. NOTICE Contact corrosion due to moisture. Damage to product.
  • Page 26 The medical device is resistant to temperatures of up to 138 ℃ (280.4 ▶ Place the cleaned and disinfected handpieces separately in sterilisation packages and weld them to be sealed, e.g. KaVo STERIclave bags Mat. no. 0.411.9912). 135°C KaVo recommends, e.g.
  • Page 27: Reprocessing Of The Assistant Element

    2 Reprocessing methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant element Note The use of hot-air sterilisation and radio-sterilisation is not permissible (causes destruction of the products). KaVo shall not be held responsible if non-permissible procedures such as ethylene oxide, formaldehyde, and low temperature plasma sterilisation are used.
  • Page 28: Reprocessing Of The Suction Hose Connector

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.5 Reprocessing of the assistant element ▶ Wipe-disinfect the instrument holder. 2.5.2 Reprocessing of the suction hose connector ▶ Pull-off the connectors ① and ② from the suction hoses.
  • Page 29: Reprocessing Of Sieve Inserts And Suction Hoses

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.6 Reprocessing of the patient element 2.5.3 Reprocessing of sieve inserts and suction hoses Note Check the sieve inserts daily and after each surgery. The sieve inserts must be replaced at the latest when the suction performance decreases (Mat.
  • Page 30: Reprocessing Of The Spittoon Bowl

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.6 Reprocessing of the patient element Cleaning and disinfection of the tumbler holder and tumbler filler ▶ Take off the tumbler holder (Mat. no. 1.008.9291). ▶ Clean the tumbler holder under running water.
  • Page 31: Manual Hygiene Functions Of The Water-Conducting System

    During normal operation of the treatment centre, the continuous germ reduc- tion function (available with water bottle kit and added disinfectant KaVo OXY- GENAL 6) ensures continuous germ reduction of the water-supplying systems. Manual rinsing and manual intensive germ reduction facilitate the rinsing and germ reduction of the water-supplying system.
  • Page 32 ▶ Fill the tumbler 3x in order to rinse and disinfect the entire water-supplying system. ▶ Allow water and KaVo OXYGENAL 6 to act for at least 30 minutes, and preferably over the weekend. ▶ After the time of exposure is elapsed, fill the tumbler 3x and rinse each dental instrument (without attachment handpieces or cannulas) for 20 seconds.
  • Page 33: Reprocessing Of The Suction System

    Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.8 Reprocessing of the suction system Intensive germ reduction during prolonged periods of non-use (> 4 weeks) During extended periods of non-use, e.g. semester breaks at university, intens- ive germ reduction must be carried out during the time of non-use.
  • Page 34 Care instructions ESTETICA E30 2 Reprocessing methods according to DIN EN ISO 17664 | 2.8 Reprocessing of the suction system 34 / 40...
  • Page 35: Replenishing And Checking The Disinfectant

    Care instructions ESTETICA E30 3 Replenishing and checking the disinfectant | 3.1 Replenishing KaVo OXYGENAL 6 when using the water bottle 3 Replenishing and checking the disinfectant 3.1 Replenishing KaVo OXYGENAL 6 when using the water bottle NOTICE Damage caused by defective water bottle.
  • Page 36 Care instructions ESTETICA E30 3 Replenishing and checking the disinfectant | 3.1 Replenishing KaVo OXYGENAL 6 when using the water bottle ▶ Remove the water bottle ① and intensive germ reduction bottle ② through a slow counterclockwise rotation. Note This allows the over-pressure to be released slowly from the water bottle. If the pressure supply to the equipment is not switched off, a permanent vent- ing sound can be heard.
  • Page 37: Filling The Intensive Germ Reduction Bottle

    Care instructions ESTETICA E30 3 Replenishing and checking the disinfectant | 3.1 Replenishing KaVo OXYGENAL 6 when using the water bottle ▶ Squeeze the KaVo OXYGENAL 6 bottle until the germ reduction liquid rises above the red marker. ð In response to over-filling, the excess quantity above the tick mark is aspir- ated automatically to ensure exact dosing.
  • Page 38: Check The Oxygenal 6 Concentration

    Care instructions ESTETICA E30 3 Replenishing and checking the disinfectant | 3.2 Check the OXYGENAL 6 concentration Note Substances remaining in the dosing head are not contamination, but rather agents that are deposited there during extended periods of non-use after condensation of the water.