Malfunction
Nothing works.
The patient chair does
not move.
Display without indic-
ator.
Operating device no
function.
Turbine making loud
running noises.
Satelec Mini LED /
KaVo Poly One does not
work.
No cold light on the
handpieces.

9 Troubleshooting

Note
In case of malfunctions of the individual handpieces (e.g. turbine, motor,
camera, Satelec, Mini LED etc.), consult the separate instructions for the use
and care of the individual instruments.
Cause
Main switch is off.
Main service fuse inter-
rupted the electric cir-
cuit.
The safety shutoff is
activated.
(LED on the control
panel flashes.)
Bus / hardware error.
Bus / hardware error.
Turbine wheel faulty.
Also refer to: Instruc-
tions for Use of the
Satelec Mini LED /
KaVo Poly One
Cold light not preselec-
ted.
The high-pressure lamp
or Multi LED on the
handpiece is defective.
Remedy
▶ Turn on the main switch.
▶ Unplug the unit from the mains.
▶ Check and replace, if required, the main ser-
vice fuse.
The main service fuse is situated next to the
master switch.
▶ For this purpose, open the bayonet closure
with a screwdriver and replace the fine-wire
fuse (T 6.3 H Mat. no. 0.223.2783).
▶ The re-close the bayonet closure with the
screwdriver.
▶ Check the safety shutoff and eliminate the
reason for the shutoff.
▶ Turn the device off and on.
▶ Call the service technician to look into the
problem if it continues to exist.
▶ Turn the device off and on.
▶ Call the service technician to look into the
problem if it continues to exist.
▶ Replace turbine wheels.
Follow the operating instructions for the tur-
bine.
▶ Preselect cold light.
▶ Replace the high-pressure lamp or Multi LED.
Also refer to: Instructions for Use of the
handpiece
See also:
2 Instructions for Use of the handpiece
125 / 134
Instructions for use ESTETICA E30
9 Troubleshooting