Instructions for use EXPERTmatic E31 C – 1.008.4998 Table of contents Table of contents 1 User instructions....................... 2 Safety ..........................2.1 Infection hazard......................7 2.2 Improper use ....................... 7 2.3 Technical condition......................7 2.4 Accessories and combination with other equipment ............8 2.5 Qualification of personnel....................
EXPERTmatic is a registered trademark of KaVo Dental GmbH. All other trademarks are property of their respective owners. KaVo Original Factory Repair In the event of a repair, please ship your product to the KaVo Original Factory Repair using www.kavobox.com. KaVo Technical Service...
Page 5
Instructions for use EXPERTmatic E31 C – 1.008.4998 1 User instructions Serial number UDI symbol Manufacturer Note: Please note accompanying documents Follow the electronic instructions for use HIBC Code CE mark for medical devices EAC conformity mark (Eurasian Conformity) Medical device, labelling of medical devices...
Page 6
Instructions for use EXPERTmatic E31 C – 1.008.4998 1 User instructions CAUTION In cases which – if not prevented – can lead to minor or moderate injury. CAUTION In cases which – if not prevented – can lead to material damage.
4 Do not operate the medical device at eye level. 2.3 Technical condition A damaged product or damaged or NOT KaVo original components could injure patients, users or third parties. 4 Only operate devices or components if they show no signs of damage on the outside.
Repairs, servicing and safety checks may only be performed by trained service personnel. The following persons are authorised to do this: ▪ Service technicians of KaVo branches after the appropriate product training ▪ Service technicians of KaVo authorised dealers after the appropriate product...
Page 9
Define the service interval depending on the frequency of use. As a result of the use of NON-KaVo original spare parts during the repair, parts may become detached and injure patients, users or third parties. This may res- ult in aspiration, swallowing of parts and possibly even a risk of suffocation.
Instructions for use EXPERTmatic E31 C – 1.008.4998 3 Description of the product | 3.1 Intended use 3 Description of the product EXPERTmatic Contra-angle handpiece E31 C (Mat. no. 1.008.4998) 3.1 Intended use Indications for use: This medical device is: ▪...
Instructions for use EXPERTmatic E31 C – 1.008.4998 3 Description of the product | 3.3 Transportation and storage conditions NOTE Operate the product with clockwise rotation. NOTE Do not operate at maximum speed. Adjust the speed via the motor control.
Dispose of and recycle the packaging appropriately in accordance with current packaging law, employing waste management / recycling companies. Comply with the comprehensive return system. KaVo has had its packaging licensed for this purpose. Please comply with the regional public waste-disposal system.
Instructions for use EXPERTmatic E31 C – 1.008.4998 5 Operation | 5.1 Attaching the medical device 5 Operation WARNING The patient swallows or aspirates loose parts or substances. Danger of suffocation. 4 Place a rubber dam before each treatment. NOTE...
Instructions for use EXPERTmatic E31 C – 1.008.4998 5 Operation | 5.4 Removing the dental bur WARNING Use of unauthorised dental burs. Injury to the patient or damage to the medical device. 4 Comply with the instructions for use and the intended use of the dental bur.
Page 15
Instructions for use EXPERTmatic E31 C – 1.008.4998 5 Operation | 5.4 Removing the dental bur 4 Unscrew it or take it off the insert after it has stopped rotating. NOTE For the re-use of the inserts, please refer to the instructions for use from the manufacturer.
Malfunction or complete failure. 4 Do not use Vaseline or other grease or oil. 4 Spray a lint-free cloth with KaVo Spray and apply it to the O-rings on the coupling. 4 Press the O-ring between your fingers to form a loop.
Page 17
Instructions for use EXPERTmatic E31 C – 1.008.4998 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Spray a lint-free cloth with KaVo Spray, apply it to the new O-rings and in- sert them in the recesses. 17 / 28...
Malfunction and material damage. 4 Reprocess in a washer disinfector or by hand only. KaVo recommends the following products based on the compatibility of the ma- terials. The microbiological efficacy must be ensured by the disinfectant manu- facturer and proven by an expert opinion.
Instructions for use EXPERTmatic E31 C – 1.008.4998 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.2 Manual Reprocessing CAUTION Never reprocess this medical device in an ultrasonic cleaner. Malfunction and material damage. 4 Reprocess in a washer disinfector or by hand only.
7.2.3 Manual drying Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts. 4 Cover the medical device with the KaVo Cleanpac bag, and place it on the corresponding care adapter. 4 Hold the can vertically.
NOTE Please comply with the instructions for use of the washer disinfector. 4 In order not to affect the KaVo medical device, make sure that the product is dry on the inside and outside after completion of the cycle. 4 Remove any residual liquids with KaVoDRYspray.
Instructions for use EXPERTmatic E31 C – 1.008.4998 7 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 7.6 Sterilisation 7.6 Sterilisation Sterilisation in a steam steriliser (autoclave) in accordance with EN 13060 / EN ISO 17665-1 CAUTION Improper service and care.
Instructions for use EXPERTmatic E31 C – 1.008.4998 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat. no. Holder 1.004.0596 INTRA Instrument stand 3.005.5204 Cleanpac 10 units 0.411.9691 Cellulose pad 100 units 0.411.9862...
24 months from the date of the invoice, subject to the following conditions: With regard to justified complaints KaVo grants warranty in the form of a free of charge repair or delivery of a replacement. Other claims of any kind whatso- ever, in particular with respect to compensation, are excluded.