Key
Settings
S3
Changes the torque setting display
Unit: % or Ncm
Transmission ratio 1:1: 100% = 2.5 Ncm
Transmission ratio 3:1: 100% = 8.0 Ncm
S4
Selects the next option
Option 3: AutoRev. / Fwd
▶ Press the "S2" or "S3" key to set the period of time after which the motor
rotates clockwise automatically. This dispenses with having to stop it using
the foot control.
S1
S2
S3
Key
Settings
S1
Saves the option
Selects the ENDO menu
S2
Sets the time
1 second to 10 seconds
S3
Sets the time
1 second to 10 seconds
S4
Selects the next option

4.8.7 Using the CONEXIOcom (optional)

Note
To start the CONEXIOcom menu, no handpiece may be removed.
Note
For all CONEXIOcom functions, the dental unit must be connected to an in-
stallation of the KaVo "CONEXIO" software.
The function of the CONEXIOcom menu is to control the display of previously
recorded and saved images and videos. In order to use the function, the unit
must have access to the data of the KaVo Software "CONEXIO" software. For
details on the configuration, please refer to the "CONEXIO" installation instruc-
tions.
Opening the CONEXIOcom menu
In order to display extant images, open the menu with the "Remote Control"
button. Select the proper patient on the corresponding PC for this purpose. It is
also feasible to automatically transfer the patient from your invoicing pro-
gramme to CONEXIO. For details on the configuration, please refer to the
"CONEXIO" installation instructions.
4 Operation | 4.8 Using functions through the menu