Malfunction
ERGOcam/DIAGNOcam
does not work.
No data transmission to
the multimedia menu
of the unit.
Camera images shows
images only as black/
white images.
Camera image freezes
without the release
button or foot control
having been triggered.
Camera image fails to
return to live image
mode.
Camera image freezes
without the release
button or foot control
having been triggered.
Taking the camera out
again did not solve the
problem.
Camera image freezes
without the release
button or foot control
having been triggered.
The monitor turns itself
off.
An acoustic signal is is-
sued every second.
Cause
PC is switched off.
USB cable too long.
No or faulty ethernet
connection between
dental unit and office
network.
Electrical or electro-
magnetic interference
by other equipment.
Electrical or electro-
magnetic interference
by other equipment.
Electrical or electro-
magnetic interference
by other equipment.
Electrical or electro-
magnetic interference
by other equipment.
Leaking water switch
recognises leaking wa-
ter.
127 / 134
Instructions for use ESTETICA E30
Remedy
▶ Turn on the computer.
▶ Make sure that the cable length does not ex-
ceed 10 m (2 x 5 m passive with repeater).
▶ Notify network administrator.
▶ Restart the CONEXIO PC.
▶ Replace the camera in the holder and then
take it out again.
▶ Restart the software.
▶ Restart the treatment unit and the CONEXIO
PC.
▶ Remove water from the unit body. If neces-
sary, have a technician fixed the leak.