7 Safety checks - testing instructions | 7.2 Instructions for safety checks
Connecting the safety tester without the KaVo
measuring cable to the treatment centre.
1
① Protective earth conductor terminal
(PE)
▶ Switch L + N of the on-site power supply cord to be voltage-free.
▶ Disconnect L + N on terminals X8.L Mains (mains supply) and X8.N Mains
(mains supply).
▶ Connect the safety tester directly to terminals X8.L Mains (mains supply)
and X8.N Mains (mains supply) and protective earth conductor terminal
(PE).
Note
The main switch of the ME device / ME system must be turned on during the
measurement.
Connect the application parts [AP] to the safety tester:
▶ Connect ① to ③ to the safety tester.
▶ Connect the safety tester to additional measuring points AP X.
Note
Additional measuring points AP X must be taken into consideration in the
presence of accessories: e.g. accessories such as PIEZO ultrasonic scaler,
etc.