KaVo ESTETICA E30 Short Instructions For Use
Hide thumbs Also See for ESTETICA E30:

Advertisement

Short instructions for use
ESTETICA E30

Advertisement

Table of Contents
loading

Summary of Contents for KaVo ESTETICA E30

  • Page 1 Short instructions for use ESTETICA E30...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH Kaltenbach & Voigt GmbH Bismarckring 39 Bismarckring 39 D-88400 Biberach D-88400 Biberach Phone +49 (0) 7351 56-0 www.kavo.com Fax +49 (0) 7351 56-1488...
  • Page 3: Table Of Contents

    Table of contents Table of contents Controls .......................  4 Dentist element TM table .................  4 Dentist element S-table ................ 4 Assistant element.................. 5 Groups of keys .................. 5 Foot control....................  7 Moving the patient chair ................ 9 Positioning the dental chair manually ............ 9 Positioning the patient chair with the foot control ........
  • Page 4: Controls

    Prior to first startup of the unit, the accompanying documents need to be inspected! The current accompanying documents are available for downloading at www.kavo.com . 1.1 Dentist element TM table Group of keys for the dental chair...
  • Page 5: Assistant Element

    Short instructions for use 1 Controls 1.3 Assistant element Group of keys for hygiene Group of keys for illumination Group of keys for the timer Group of keys for the dental chair 1.4 Groups of keys Group of keys for the dental chair The keys of the assistant unit each have two functions and show two symbols.
  • Page 6 Short instructions for use 1 Controls Assistant element key Dentist element key Name "AP 2" key (automatic position 2) "Collapsed position" key Group of keys for illumination/handpieces Name Control element Dentist element and assist- "Operating light" ant element "Operating light dimming" Dentist element Dentist element "X-ray viewer"...
  • Page 7: Foot Control

    Short instructions for use 1 Controls Group of keys for the menu Group of keys for the menu ① ② Menu function selection keys Display Group of keys for the timer Description Control element "Remote Control" (CON- unassigned key EXIO) key "Additional motor drives"...
  • Page 8 Short instructions for use 1 Controls Item Name with mounted with removed Handpiece Handpiece "Preselection of level/ Preselection of level Starts the handpieces and ⑤ handpieces" foot-pedal controls the speed/in- tensity of the handpieces. "Chair position" cross switch ① Chair up ②...
  • Page 9: Moving The Patient Chair

    Short instructions for use 2 Moving the patient chair 2 Moving the patient chair 2.1 Positioning the dental chair manually Function The chair moves up. The chair moves down. The backrest moves upward. The backrest moves downward. 2.2 Positioning the patient chair with the foot control See also: 2 "Chair position"...
  • Page 10 Short instructions for use 2 Moving the patient chair ▶ During these four seconds, briefly press the "AP 0", " AP 1", " AP 2", "SP" or "LP" key. ð The chair moves into the selected automatic position. Saving an automatic position through the assistant element ▶...
  • Page 11: User Menu

    Short instructions for use 3 User menu 3 User menu The following options can be opened in the user menu: Option Function Description Users Set number of users. Tumbler Set tumbler filling time. Bowl Set bowl rinsing time. Handpiece light Set cold light afterglow period.
  • Page 12: Selecting And Setting An Option

    ▶ Press the "S1", "S2", "S3" key to select user 4, 5, and 6, respectively. 3.2 Selecting and setting an option ▶ Press the "Increase value" and "Decrease value" keys to select the de- sired parameter or value. See also: 2 Instructions for use ESTETICA E30...
  • Page 13: Standby Menu

    Short instructions for use 4 Standby menu 4 Standby menu The stand-by menu is used to start the device. The device switches to the stand-by menu upon closing of the handpiece and endo menus. ▶ Press the "Next" key (S4) to start-up the user menu. ð...
  • Page 14: Setting The Timer

    Short instructions for use 5 Setting the timer 5 Setting the timer 5.1 Select the timer time ▶ Press the "Timer" key briefly to start or stop the timer. ð LED flashes while the timer counts down. 5.2 Set the timer time ▶...
  • Page 15: Setting The Handpieces

    Short instructions for use 6 Setting the handpieces 6 Setting the handpieces 6.1 Adjusting functions Settings can be changed using the corresponding keys on the control element or on the foot control. The motor speed can be set with the foot control only. The direction of motor rotation, cooling status and light intensity can be set with the foot control or keys on the control element.
  • Page 16: Turbine

    Short instructions for use 6 Setting the handpieces Setting the direction of motor rotation Note The direction of motor rotation can only be changed when the motor is at rest. ▶ Take motor off the holder. ð The handpiece is active. ▶...
  • Page 17 Short instructions for use 6 Setting the handpieces Setting the PIEZOsoft Settings can be changed using the corresponding keys on the control element or on the foot control. The intensity can be set with the foot control only. The spray status can be changed either with the foot control or the key on the control element.
  • Page 18: Kl 703 Led In Endo Mode (Optional Accessories)

    Short instructions for use 6 Setting the handpieces 6.5 KL 703 LED in ENDO mode (optional accessories) Note The functions of the ENDO menu are available only if an ENDO license is existent. Start-up ENDO menu ▶ Take the ENDO motor out of the dentist element (setting in user menu, option 5). ▶...
  • Page 19 Short instructions for use 6 Setting the handpieces Parameters Values Transmission factor 1:1, 3:1 Parameter memory P1,P2,P3,P4,P5,P6 Torque mode TQ-Ctrl, Autoreverse, AutoRev/Forw. Direction of motor rotation R, L Speed 100 to 6,000 rpm Torque Transmission ratio 1:1: 0.15 to 2.50 Ncm Transmission ratio 3:1: 0.4 to 8.0 Ncm The following parameters can be changed: •...
  • Page 20: Operate Handpieces With The Foot Control

    Short instructions for use 7 Operate handpieces with the foot control 7 Operate handpieces with the foot control ▶ Remove the handpiece (e.g turbine, motor) from the holder. ð The handpiece is active. ▶ Press the foot pedal. ð The removed handpiece runs at the set speed or intensity. ▶...
  • Page 21: Hygiene

    During normal operation of the treatment centre, the continuous germ reduction function (available with water bottle kit and added disinfectant KaVo OXYGENAL 6) ensures con- tinuous germ reduction of the water-supplying systems.
  • Page 22: Reprocessing Of The Suction System

    ▶ Fill the tumbler 3x in order to rinse and disinfect the entire water-supplying system. ▶ Allow water and KaVo OXYGENAL 6 to act for at least 30 minutes, and preferably over the weekend. ▶ After the time of exposure is elapsed, fill the tumbler 3x and rinse each dental instru- ment (without attachment handpieces or cannulas) for 20 seconds.
  • Page 23 Short instructions for use 8 Hygiene ▶ With the suction handpiece attached or by hand, press down the DEKASEPTOL dosing facility once each for the saliva ejector hose and the spray mist ejector. ▶ Aspirate the DEKASEPTOL Gel from the dosing facility. Note In suction processes with air jet suction systems (Venturi), the two fol- lowing steps must not be carried out as they could lead to blockage of...
  • Page 24: Replenishing Oxygenal

    ▶ Remove the protective hose from the tube of the KaVo OXYGENAL 6 dosing facility. ▶ Guide the tube through the orifice of the coarse sieve of the KaVo OXYGENAL 6 bottle. ▶ Screw the KaVo OXYGENAL 6 dosing facility tightly to the KaVo OXYGENAL 6 bottle.
  • Page 25 ▶ Use KaVo OXYGENAL 6 only. ▶ Do not exceed the maximum concentration of 0.25% KaVo OXYGENAL ▶ Proceed like for filling the water bottle, but use a different dose of KaVo OXYGENAL 6. See also: 2 Filling the water bottle ▶...
  • Page 26 Short instructions for use 8 Hygiene ▶ Then fill-up the intensive germ reduction bottle with tap water. ▶ Screw the intensive germ reduction bottle tightly to the treatment centre through a clockwise rotation. ð Apply compressed air to the intensive germ reduction bottle. Check the OXYGENAL 6 concentration ▶...
  • Page 27: Troubleshooting / Warning Signals

    Short instructions for use 9 Troubleshooting / warning signals 9 Troubleshooting / warning signals 9.1 Safety shut-off Pos. Safety switch-off actu- LED on assistant ele- LED on dentist element ated ment Bracket on the foot con- ① trol Assistant element ②...
  • Page 28: Troubleshooting

    Satelec Mini LED / Also refer to: In- KaVo Poly One does structions for Use of not work. the Satelec Mini LED / KaVo Poly One No cold light on the Cold light not ▶ Preselect cold light. handpieces. preselected.
  • Page 29 Short instructions for use 9 Troubleshooting / warning signals Malfunction Cause Remedy No spray in the No spray preselec- ▶ Preselect spray. Check setting on butterfly valves of the handpieces. ted. dentist element. Close the ring for ▶ Open the ring for controlling the spray on the handpieces.
  • Page 30: Service And Error Messages In The Stand-By Menu

    Short instructions for use 9 Troubleshooting / warning signals Malfunction Cause Remedy The suction hoses do Slides on the conical ▶ Open the slide valve. not have any suction. sections are closed. Sieves in suction ▶ Replace sieves. connector are blocked.
  • Page 31 Short instructions for use 9 Troubleshooting / warning signals Error messages in the status display Malfunction Cause Remedy Display shows: ID 64 Water is shut off. ▶ Turn water on. Water works leaks ▶ Call a technician. strongly. Water works mal- function Display shows: ID 65 Safety switch of bowl ▶...
  • Page 32 Short instructions for use 9 Troubleshooting / warning signals...