3 Product description | 3.10 Technical data
Cannula connecting pieces
Diameter of cannula of saliva ejector
and surgical suction
Diameter of cannula of spray mist
suction
Suction air quantity at spray mist
cannula
Diameter of the suction connection
Suction connection above floor
Operating environment
Inappropriate operating conditions.
Impairment of the electrical safety of the device.
▶ It is essential to comply with the operating conditions specified in the
"Technical Specifications" chapter.
Floor quality
Ambient temperature
Relative humidity
Air pressure
Max. elevation for operation
Maximum loads
Max. patient weight load
Tray holder of the dentist element -
loadable up to
Assistant element standard tray -
loadable up to
Dentist element - loadable up to
Transportation and storage conditions
Ambient temperature
Relative Humidity
Air pressure
according to DIN EN ISO 7494 - 2
7 mm
15 mm
minimum 250Nl/min (suction system
with high flow rate), recommended
300Nl/min
40 mm
20 mm
WARNING
The quality of the flooring must meet
the load bearing ability for buildings
DIN 1055 page 3 and have a pres-
sure resistance in accordance with
DIN 18560 T 1.
+10 to +40
30 to 75%
700 hPa - 1,060 hPa
up to 3000 m
180 kg
2 kg
1 kg
2 kg
-20 to +55
5% to 95% non-condensing
700 hPa to 1,060 hPa
38 / 134
o
C
o
C