7 Safety checks - testing instructions | 7.2 Instructions for safety checks
See also:
2 8 Annex - Additional measuring points, Page 122
Connect accessible conductive parts [ACP] with PE
ACP = accessible conductive parts
ACP
Note
Additional measuring points ACP X must be taken into consideration in the
presence of accessories.
See also:
2 8 Annex - Additional measuring points, Page 122
ACPs on the treatment centre
No ACPs need to be connected to the protective conductor (PE) during the
measurement on the treatment centre ESTETICA E30, as all relevant parts are
connected to the PE at the factory and are included in the test.
ACPs on treatment lamps
No ACPs need to be connected to the operating lights during the measurement
with protective conductor (PE) because all relevant parts have already been
connected to the protective conductor (PE) at the factory and are included in
the test.