4 Operation | 4.12 Using the KL 703 LED in ENDO mode (optional accessory)
▶ To adjust the intensity, press the foot pedal and move it to the side.
4.12 Using the KL 703 LED in ENDO mode (optional
accessory)

4.12.1 General information

Note
The endo drive can only be operated with the INTRA LUX KL 703 LED.
INTRA LUX KL 703 LED
Use of impermissible filing systems.
Do not use impermissible filing systems which can damage the product or
cause personal injury.
▶ Only use approved NiTi filing system with a conicity >2% that are suitable
for rotary preparation.
▶ Only use files with shafts in conformance with DIN EN ISO 1797-1, DIN EN
ISO 1797-2, DIN EN ISO 3630-1 and DIN EN ISO 3630-2 having a shaft
diameter of 2.334 to 2.350 mm
▶ Follow manufacturer's instructions (mode of operation, speed, torque
levels, torsion resistance, etc.), and use the files according to their intended
use.
Use of damaged files.
Injury to the patient or damage to the medical device.
▶ Before preparing each root canal, insert a dental dam for safety reasons.
▶ Before each use, the files must be checked for possible signs of
material fatigue, deformation or excessive stress and if such signs appear,
they must be replaced.
Incorrect transmission ratio.
Damage from incorrect speed / incorrect torque.
▶ Only use KaVo 1:1 reducing shanks 20LH or 20LP or MASTERmatic LUX
M20 L with 1:1 INTRA LUX head L68 B (Mat. no. 1.008.1834) or 3:1 IN-
TRA head L66 B (Mat. no. 1.008.1831).
CAUTION
CAUTION
CAUTION
93 / 134
Instructions for use ESTETICA E30