7 Safety checks - testing instructions | 7.2 Instructions for safety checks
*The EUL limit corresponds to the value defined in IEC 60601 (DIN EN 60601),
taking comment 2 from table 2 into consideration.
7.1.6 Notes on repeat testing
Note
The value determined in these tests must be documented and evaluated to-
gether with the measuring process. The measured values must not exceed
the specified values.
Note
Comparisons with previous measurements must be carried out if the meas-
ured values are lower than the threshold values by more than 10 %. The test
intervals should be reduced if a deterioration in values is determined!
7.2 Instructions for safety checks
7.2.1 Preparatory measures to be undertaken on the
device
Electrical power.
Death or injury from electric shock.
▶ Before servicing, pull the mains plug out of the socket or completely discon-
nect the device from the power to de-energise it!
▶ After conversion, check the electrical safety in accordance with IEC 62353
(DIN VDE 0751-1).
▶ Turn off the main switch before any maintenance work.
▶ Lift the upholstery at the foot end ① and remove from the chair.
▶ Remove two screws ② from the cover of the junction box.
WARNING
105 / 134
Instructions for use ESTETICA E30