Technical Data - KaVo ESTETICA E30 Instructions For Use Manual

Hide thumbs Also See for ESTETICA E30:
Table of Contents

Advertisement

Instructions for use ESTETICA E30
3 Product description | 3.10 Technical data

3.10 Technical data

Drilling template and setup plan
Layout plan (Mat. no. 1.009.2781)
Electrical system
Electrical lead
Free end above the floor
Input voltages
Frequency
Factory-set mains input voltage
Power consumption
Customer-provided fuse protection
Protective conductor above floor
Heat emission
Heat emission
Degree of soiling
Installation category
Mark of approval
Foot control
Triple function handpiece One
Flush the water and air passages for 20 to 30 seconds before working at the
beginning of the day.
Water pressure
Water flow
Air pressure
Air flow
Triple function handpiece
Flush the water and air passages for 20 to 30 seconds before working at the
beginning of the day.
3 x 1.5 mm
fuse protection 10 A) 3 x 2.5 mm
(customer-provided fuse protection
16 A)
1 000 mm
100/110/115/120/127/220/230/240
V AC
50/60 Hz
See nameplate
100 to 900 VA
Auto-mat C16 or screw-plug fuse 10
A
See DIN VDE 0100-710, 1000 mm
360 to 3240 kJ/h
Ø 900 kJ/h
2
II
CE / DVGW / VDE
IPX1: Protection against water drips
1.5 ± 0.3 bar; flow pressure; probe
tip (1.009.6634)
90 ± 15 ml/min
1.5 ± 0.2 bar; flow pressure; probe
tip (1.009.6634)
15 ± 3 Nl/min
34 / 134
2
(customer-provided
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents