4.3.5 Safety shut-off

To prevent collisions arising from the movement of the patient chair, safety
shutoff switches are installed to protect the patient and practice personnel from
injury and the treatment centre from damage.
Damage to the assistant element and dental chair.
Despite some safety shut-downs being present, certain positions of the
assistant element may collide with the dental chair.
▶ Keep the assistant element out of the motion range of the patient chair.
▶ Always monitor the chair movement.
Pinching from the treatment chair.
The safety shutoff of the treatment chair is activated by lifting the respective
component. Depending on the patient's body weight and the leverage, more
force can be exerted on the object to be triggered than is necessary to trigger
the switching function.
▶ The treatment personnel must move outside of the chair's swivelling range
whenever the chair moves.
Risk of injury when moving the patient or patient chair.
The patient or treatment personnel can be pinched or crushed.
▶ Position all moving parts, such as dentist element, assistant element, oper-
ating light, screens, etc., outside the collision range when you move the pa-
tient or patient chair.
The safety cut-offs can be found at the following places on the treatment unit.
NOTICE
CAUTION
CAUTION