4 Operation | 4.8 Using functions through the menu
S1
S2
S3
Option 1: Transmission ratio
▶ Press the "S2" or "S3" key to set the transmission factor.
S1
S2
S3
Key
Settings
S1
Saves the option
Selects the ENDO menu
S2
Changes the transmission factor
1:1 or 3:1
S3
Changes the transmission factor
1:1 or 3:1
S4
Selects the next option
Option 2: Torque
▶ Press the "S2" or "S3" key to change the torque setting display.
S1
S2
S3
Key
Settings
S1
Saves the option
Selects the ENDO menu
S2
Changes the torque setting display
Unit: % or Ncm
Transmission ratio 1:1: 100% = 2.5 Ncm
Transmission ratio 3:1: 100% = 8.0 Ncm
S4
S4
S4
72 / 134