4 Operation | 4.9 Using function through the dentist or assistant unit
▶ Press the "Bowl flush" key for more than 4 seconds to start the program-
ming mode.
Set the period of time in 200 ms increments. Minimum: 0.4 s.
ð
If the key stays depressed, the elapsed time is counted in 200 ms steps and
an acoustical signal is issued each second.
ð
Once the key is released, the current value is saved.
4.9.2 Using the illumination functions for KaVoLUX 540
LED T
Unintentional activation of the KaVo KEY Laser III and KEY Laser 3+.
Simultaneous application of the operating light KaVoLUX 540 LED and the
KaVo KEY Laser III or KEY Laser 3+ can lead to the unintentional activation of
the KaVo KEY Laser III and KEY Laser 3+.
▶ When using the KaVo KEY Laser III or the KEY Laser 3+, switch the operat-
ing light to laser mode.
▶ Or switch off the operating light, do not use the KaVo KEY Laser III or KEY
Laser 3+ and the KaVoLUX 540 LED operating light simultaneously.
Wrong handling.
Reversible blinding (temporary sight impairment).
▶ Do not direct the light field at patients, users or/and third parties while you
switch on the light.
▶ Do not direct the light field at the patient´s eyes when you move the light
head.
▶ Maintain a clearance of ca. 700 mm between the light and the mouth of the
patient.
Stroboscopic effect of the rotating instrument.
A stroboscopic effect could arise in instruments rotating at a certain speed dur-
ing application of the KaVoLUX 540 LED. This is an optical illusion, where the
instrument appears to be standing still or rotating extremely slowly.
Injury hazard.
▶ If the stroboscopic effect appears, change the speed fractionally and con-
tinue operating in the usual manner.
Faulty measurement in connection with KaVo DIAGNOdent.
Simultaneous application of the operating light KaVoLUX 540 LED and the
KaVo DIAGNOdent can lead to faulty measurements.
▶ Switch the operating light to laser mode when using the KaVo DIAGNO-
dent.
▶ Or switch off the operating light, do not use KaVo DIAGNOdent and operat-
ing light KaVoLUX 540 LED simultaneously.
Premature hardening of composite fillings.
A light intensity that is too high can have a negative impact on the durability of
the treatment.
▶ Select the appropriate dimming level according to the processing time.
WARNING
WARNING
CAUTION
CAUTION
CAUTION
77 / 134
Instructions for use ESTETICA E30