10 Information about electromagnetic compatibility in accordance with EN60601-1-2 | 10.1 Electromagnetic Trans-
missions
Measurements of emitted in-
terference
HF emissions according to CISPR
11
HF emissions according to CISPR
11
Emission of harmonics according
to EN 61000-3-2
Emission of voltage fluctuations/
flicker according to EN 61000-3-3
Interference im-
munity tests
Electrostatic discharge
(ESD) according to EN
61000-4-2
10 Information about electromagnetic compatibility in
accordance with EN60601-1-2

10.1 Electromagnetic Transmissions

The treatment unitESTETICA E30 is designed for operation in an environment
as specified below. The customer or user of the ESTETICA E30 should make
sure that the device is used in an environment of the specified type.
Conformance
Group 1
Class B
Class A
Conforms

10.2 Resistance to electromagnetic interference

The treatment unitESTETICA E30 is designed for operation in an environment
as specified below. The customer or user of the ESTETICA E30 should make
sure that the device is used in an environment of the specified type.
EN 60601 test level
± 6 kV contact discharge
± 8 kV atmospheric dis-
charge
Compliance level
± 2/4/6 kV contact dis-
charge
± 2/4/8 kV atmospheric
discharge
128 / 134
Electromagnetic environment
-
Guidelines
The ESTETICA E30 device uses
HF energy for its internal func-
tions exclusively. Therefore, the
HF emission of the device is very
low and interference with adja-
cent electronic devices is unlikely.
The ESTETICA E30 device is suit-
able for use in all facilities includ-
ing residential ones, and facilities
that are directly connected to a
public power supply that also
supplies residential buildings.
The ESTETICA E30 device is suit-
able for use in all facilities includ-
ing residential ones, and facilities
that are directly connected to a
public power supply that also
supplies residential buildings.
The ESTETICA E30 device is suit-
able for use in all facilities includ-
ing residential ones, and facilities
that are directly connected to a
public power supply that also
supplies residential buildings.
Electromagnetic en-
vironment -
Guidelines
Floors should be made of
wood or concrete or be
fitted with ceramic tiles.
If the floor is fitted with
synthetic material, the
relative humidity must
be at least 30%.