Instructions for use MASTERtorque Mini LUX M8700 L - 3.001.0000MASTERtorque Mini LUX M8700 LK -
3.001.3400MASTERtorque Mini LUX M8700 LS - 3.001.0600
6 Troubleshooting | 6.1 Changing the O-rings on the MULTIflex (LUX) / MULTIflex LED coupling, M8700 L
6 Troubleshooting
6.1 Changing the O-rings on the MULTIflex (LUX) /
MULTIflex LED coupling, M8700 L
Improper care of the O-rings.
Malfunction or complete failure.
▶ Do not use Vaseline or other grease or oil.
Note
The O-rings on the coupling may only be lubricated with a cotton ball wetted
with KaVo Spray.
▶ Press the O-ring between your fingers to form a loop.
▶ Push the O-ring to the front, and remove it.
▶ Insert new O-rings into the grooves.
6.2 Cleaning the spray nozzle
Hazard from nonsterile products.
Infection danger to the care provider and patient.
▶ Reprocess and sterilise the medical device properly before the next use.
Hazard from insufficient spray water.
Overheating of the medical device and damage to the tooth.
▶ Check the spray water channels and clean the spray nozzles with the nozzle
needle Mat. no. 0.410.0921 if necessary.
▶ Check the water filter and exchange if necessary.
6.3 Changing the water filter
Hazard from nonsterile products.
Infection danger to the care provider and patient.
▶ Reprocess and sterilise the medical device properly before the next use.
NOTICE
WARNING
CAUTION
WARNING
16 / 30