Errors Common to All Communication States

These errors occur in common regardless of communication state. When these errors occur, the ALARM
output of parallel I/O turns ON, and "SYSERR" and error code "CODE.XX" are displayed on the main and sub-
displays, respectively.
Error Code
Name
02
Linearity correction
data error
03
Linearity correction
data read error
07
System setting error
08
Bank data error
11
Firmware update
error
25
Ethernet
communication
parameter error
If an error code other than the one listed above is displayed, the displacement sensor may be broken. Please
contact an OMRON branch or sales office.
*1:
As a provisional measure, the measurement can be resumed using the data of the previously read Calibration ROM.
<Operation method>
With error code 3 displayed on the sub-display, hold down the Mode switching key, then when [OK/CAN] is displayed, press the ZERO/SET
key.
<Cautions>
- When using this method, always check the serial number of the previously read Calibration ROM in "controller information," then check
that it matches the Sensor Head side serial number. Measurement will not be correct unless they match.
- When restarting the main unit, perform the same operations again.
This operation is disabled for a displacement sensor into which no Calibration ROM has ever been inserted and started up.
ZW User's Manual
Description
Cause
The linearity correction
Calibration ROM damage Check to make sure that the
data is corrupted.
Reading of the linearity
Calibration ROM not
correction data was not
inserted
executed correctly.
The system settings
The displacement sensor
saved to the Sensor
power was turned OFF
Controller are corrupt.
during saving/loading of
system settings.
The bank data saved to
The displacement sensor
the Sensor Controller is
power was turned OFF
corrupt.
during saving/loading of
bank data.
Failed to update the
WarpEngineZW was
firmware by
interrupted during an
WarpEngineZW.
update.
An invalid IP address is
Invalid IP address setting Change to the correct IP address.
set.
Remedy
Calibration ROM is correctly
inserted. If correctly inserted, the
Calibration ROM or displacement
sensor may be broken.
Please contact an OMRON branch
(*1)
or sales office.
After holding down the Mode
switching key, press the ZERO/SET
key to clear the system settings and
the bank data, then resume the
starting process.
Do not operate WarpEngineZW
during an update.
If this error occurs, turn the
displacement sensor ON again and
retry the update.
Error Messages
209