8-1 Error Messages

Errors for Ethernet Connection

The error log for the following errors that occur in Ethernet communications can be checked on the digital
displays.
Error Code
Name
03D0 Hex
Ethernet
communication
ALARM
parameter error
03D3 Hex
Ethernet link not
detected
Note When error marked by
main and sub-displays, respectively.
If an error code other than the one listed above is displayed, the displacement sensor may be broken. Please
contact an OMRON branch or sales office.
208
Error Messages
Description
An invalid IP address is
set.
The Ethernet link cannot
be detected.
ALARM
occur, the ALARM output of parallel I/O turns ON, and "SYSERR" and error code are displayed on the
Cause
Remedy
Invalid IP address
Change to the correct IP address.
setting
Link with switching
Inspect the following items:
hub not detected
• Are cables connected?
• Are cables disconnected or loose?
• Is there a lot of noise?
ZW User's Manual

Advertisement

Table of Contents

Troubleshooting