Omron ZW-C1*T User Manual page 79

Displacement sensor zw series confocal fiber type
Hide thumbs Also See for ZW-C1*T:
Table of Contents

Advertisement

Automatically Following the Range of Measurement Area2
When 2 area mode is set, the measurement area 2 can be set automatically to measure the measurement
surface (Follow) , and the measurement area 2 range can be follow the measurement surface (Base)
movement.
This function is used when a few surface of vibrating object is measured stabile.
Item
Setting item
Area follow
Area follow
settings
Base
Follow
 Explorer pane
 Edit pane
 Sensing settings window : [Measurement area 2] - [Area follow setting]
ZW User's Manual
Base
Follow
Setting value
Description
OFF (default
Area follow is not set. Select this to set any desired value.
value)
Upper line
Adjust the measurement area of measurement area 2 to track the [Start]
setting value for measurement area 1.
Lower line
Adjust the measurement area of measurement area 2 to track the [End]
setting value of measurement area 1.
Upper line and
Adjust the measurement area of measurement area 2 to track the [Start] and
lower line
[End] setting value of measurement area 1.
Edge1/Edge2/
Select the measurement surface of measurement area 1 to be used as the
Edge3/Edge4/
reference for automatic tracking.
peak
The peak is the surface of all measurement surfaces having the highest
(default value:
received light amount.
Edge1)
Edge1/Edge2/
Select the measurement surface of measurement area 1 to be used as the
Edge3/Edge4/
reference for automatic tracking.
peak
The measurement area 2 is automatically set referenced to the surface
(default value:
selected when [Area Teach] is clicked.
Edge2)
The peak is the surface of all measurement surfaces having the highest
received light amount.
: [Bank] | [(Bank Data Name)] (double click)
: [Sensing setting] icon (
Area2 upper line
Area2 lower line
)
77