5.3

Configuring the Rosemount 5708L

5.3.1

Basic Settings

Configuration - Rosemount 5708L
<tag name>
3.45m Avg Dist.
█████████████▒▒▒▒▒
Menu
Basic Settings
Advanced Settings
False Echo Mapping
Polling Address
m
ft
Set Vessel Height
20.000 m
Cylindrical Vessel
Rectangular Vessel
Set Vessel Diameter
10.000 m
Set Scanner Height
20.000 m
Configuration - Rosemount 5708L
1. From the main screen, select the
2. In the main menu, scroll down to Basic Settings, using the
Select
to switch to the Basic Settings configuration screen.
E
3. Set the distance units, either meters (m) or feet (ft).
4. Set Vessel Height from the vessel bottom edge to the vessel top
edge.
Use the
key to switch between the two digits. Use the
modify the value.
5. Select the vessel shape, either Cylindrical or Rectangular.
a. Use the
/
keys to switch between the options. Select
to continue.
b. If Cylindrical is selected, set vessel diameter.
Use the
key to switch between the two digits. Use the
key to modify the value. Select
c. Set Scanner Height from the bottom edge of the vessel to the
mounting plate (point above the antenna).
Use the
key to switch between the two digits. Use the
key to modify the value. Select
November 2017
key to enter the main menu.
E
key to
to continue.
E
to continue.
E
key.
E
35