12. Configure the vessel full and empty calibration levels.
calibration fields to set.
It is possible to set the full and empty calibration as measured from the top (Distance) of the vessel or
measured from the bottom (Level) of the vessel.
13. Select Finish.
Figure 7-12. Configuration Wizard Step 4 - Full and Empty Calibration Levels
Figure 7-13. Full and Empty Calibration Levels Summary Table
14. When the configuration wizard settings are completed, select the Finish button. The parameters will
be uploaded to all devices in the vessel.
Operation
Set the distance from the top of the vessel to set 100% (20 mA) (see
the Full Calibration section.
OR
Set the maximal material level in the Level (Bottom) field. Dead band is 1.64 ft (0.5 m) from the
top of the Rosemount 5708 antenna case.
Set the Empty Calibration by setting the level or the distance from the top of the vessel to set the
0% (4mA) (see
Figure
7-12). Default value is set to the bottom of the vessel.
Select the Default button to set the full and empty calibration levels to default values.
Select View Table to open a window showing a summary table of each Rosemount 5708 in the
vessel and the actual full and empty calibration levels configured, see
Figure 7-12
shows the full and empty
Figure 7-12 on page
Figure 7-13 on page
Operation