Configuration - Rosemount 5708 in a System
November 2017
8.3

System requirements

Figure 8-3. Multiple Rosemount 5708 Connection
A
A. RS-485 converter
B. RS-485 (+)
C. RS-485 (-)
D. Device 1
All devices must be daisy-chained with a power supply and RS-485 links to the Rosemount System
Controller. A single 4-20mA pair must be wired from only one of the chained devices to the
PLC/DCS/SCADA/Etc. The 4-20mA may be connected to any of the devices.
The Rosemount System Controller cabinet must be installed indoors, mounted on a wall, and wired to
the devices and a power supply for the computer operation. A single 24 Vdc power supply is
recommended for powering the computer and the devices. Consider power consumption as well. The
computer is rated 40 W with 24 Vdc power supply, and each device is rated 1.5 W with the same power
supply.
140
D
E
B
H
C
F
E. Device 2
F. Device 3
G. Device n
H. 120 Ohm
Configuration - Rosemount 5708 in a System
Reference Manual
00809-0100-4570, Rev DA
G
C
H
B