Advertisement

Configuration - Rosemount 5708L
November 2017
5.3.2

Advanced Settings

After configuring the Basic Settings, perform Advanced Settings configuration.
36
Scanner Distance
From Center
00.000 m
Set Vessel Width
010.000 m
Set Vessel Length
010.000 m
Set Scanner Height
20.000 m
Scanner X To Center
±000.00 m
Scanner Y To Center
±000.00 m
<tag name>
3.45m Avg Dist.
█████████████▒▒▒▒▒
Menu
d. Set Scanner Distance from Center.
Use the
key to switch between the two digits. Use the
key to modify the value. Select
6. If Rectangular is selected, set vessel width first (dimension on the
x-axis).
a. Use the
key to switch between the two digits. Use the
key to modify the value. Select
b. Set Vessel Length (dimension on the y-axis).
Use the
key to switch between the two digits. Use the
key to modify the value. Select
c. Set Scanner Height from the bottom edge of the vessel to the
mounting plate (point above the antenna).
Use the
key to switch between the two digits. Use the
key to modify the value. Select
d. Set the distance from the x-axis, see
Use the
key to switch between the two digits. Use the
key to modify the value. Select
e. Set the distance from the y-axis, see
Use the
key to switch between the two digits. Use the
key to modify the value. Select
1. From the main screen, select the
Reference Manual
00809-0100-4570, Rev DA
to exit to the main screen.
E
to continue.
E
to continue.
E
to continue.
E
Figure 3-7 on page
16.
to continue.
E
Figure 3-7 on page
16.
to exit to the main screen.
E
key to enter the main menu.
E
Configuration - Rosemount 5708L