Sensor N Lower Limit; Sensor N Upper Limit; Sensor Serial Number - Endress+Hauser iTEMP TMT162 Operating Instructions Manual

Temperature field transmitter
Hide thumbs Also See for iTEMP TMT162:
Table of Contents

Advertisement

Operating menu and parameter description

Sensor n lower limit

Navigation
Description

Sensor n upper limit

Navigation
Description

Sensor serial number

Navigation
Description
User entry
Factory setting
94
Expert → Sensor → Sensor n → Sensor n lower limit
Displays the minimum physical full scale value.
Expert → Sensor → Sensor n → Sensor n upper limit
Displays the maximum physical full scale value.
Expert → Sensor → Sensor n → Serial no. sensor
Use this function to enter the serial number of the connected sensor.
String with up to 12 characters consisting of numbers and/or text
" " (no text)
"Sensor trimming" submenu
Sensor error adjustment (sensor trimming)
Sensor trimming is used to adapt the actual sensor signal to the linearization of the
selected sensor type stored in the transmitter. Compared to sensor transmitter matching,
sensor trimming only takes place at the start and end value and does not achieve the same
level of accuracy.
Sensor trimming does not adapt the measuring range. It is used to adapt the sensor
signal to the linearization stored in the transmitter.
Procedure
2. Set the Sensor trimming parameter to the Customer-specific setting.
3. Using a water/oil bath, bring the sensor connected to the transmitter to a known and stable temperature. A
temperature which is close to the set start of the measuring range is recommended.
1. Start
ß
ß
ß
iTEMP TMT162
Endress+Hauser

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents