Operating menu and parameter description
User entry
Factory setting
Connection type n
Navigation
Prerequisite:
Description
User entry
Factory setting
2-wire compensation n
Navigation
Prerequisite:
Description
User entry
Factory setting
Reference junction n
Navigation
Prerequisite:
70
A list of all the possible sensor types is provided in the "Technical data" section.→  46
Sensor type 1: Pt100 IEC751
Sensor type 2: No sensor
Setup → Connection type n
Expert → Sensor → Sensor n → Connection type n
An RTD sensor must be specified as the sensor type.
Use this function to select the connection type for the sensor.
• Sensor 1 (connection type 1): 2-wire, 3-wire, 4-wire
• Sensor 2 (connection type 2): 2-wire, 3-wire
• Sensor 1 (connection type 1): 4-wire
• Sensor 2 (connection type 2): none
Setup → 2-wire compensation n
Expert → Sensor → Sensor n → 2-wire compensation n
An RTD sensor with a 2-wire connection type must be specified as the sensor type.
Use this function to specify the resistance value for two-wire compensation in RTDs.
0 to 30 Ohm
0
Setup → Reference junction n
Expert → Sensor → Sensor n → Reference junction n
A thermocouple (TC) sensor must be selected as the sensor type.
iTEMP TMT162