Endress+Hauser iTEMP TMT162 Operating Instructions Manual page 51

Temperature field transmitter
Hide thumbs Also See for iTEMP TMT162:
Table of Contents

Advertisement

iTEMP TMT162
Maximum measured error
Typical
Standard
Resistance thermometer (RTD) as per standard
IEC 60751:2008
IEC 60751:2008
GOST 6651-94
Thermocouples (TC) as per standard
IEC 60584, Part 1
IEC 60584, Part 1
GOST R8.8585-2001
1)
Measured value transmitted via HART
Measured error for resistance thermometers (RTD) and resistance transmitters
Standard
Designation
Pt100 (1)
Pt200 (2)
IEC 60751:2008
Pt500 (3)
Pt1000 (4)
JIS C1604:1984
Pt100 (5)
Pt50 (8)
GOST 6651-94
Pt100 (9)
Ni100 (6)
DIN 43760 IPTS-68
Ni120 (7)
Cu50 (10)
Cu100 (11)
OIML R84: 2003 /
GOST 6651-2009
Ni100 (12)
Ni120 (13)
OIML R84: 2003, GOST
Cu50 (14)
6651-94
Endress+Hauser
In accordance with DIN EN 60770 and the reference conditions specified above. The
measured error data correspond to ±2 s (Gaussian distribution), i.e. 95.45%. The data
include non-linearities and repeatability.
Designation
Pt100 (1)
Pt1000 (4)
Pt100 (9)
Type K (NiCr-Ni) (36)
Type S (PtRh10-Pt) (39)
0 to +800 °C (32 to +1 472 °F)
Type L (NiCr-CuNi) (43)
®
.
Measuring range
–200 to +850 °C
(–328 to +1 562 °F)
–200 to +500 °C (–328 to +932 °F)
–200 to +250 °C (–328 to +482 °F)
–200 to +510 °C (–328 to +950 °F)
–185 to +1 100 °C
(–301 to +2 012 °F)
–200 to +850 °C
(–328 to +1 562 °F)
–60 to +250 °C (–76 to +482 °F)
–180 to +200 °C (–292 to +392 °F)
–180 to +200 °C (–292 to +392 °F)
–60 to +180 °C (–76 to +356 °F)
–50 to +200 °C (–58 to +392 °F)
Measuring range
0 to +200 °C (32 to +392 °F)
Maximum
≤ 0.11 °C (0.2 °F)
≤ 0.18 °C (0.32 °F)
≤ 0.11 °C (0.2 °F)
≤ 0.07 °C (0.13 °F)
≤ 0.09 °C (0.16 °F)
≤ 0.20 °C (0.36 °F)
≤ 0.11 °C (0.2 °F)
≤ 0.05 °C (0.09 °F)
≤ 0.11 °C (0.2 °F)
≤ 0.06 °C (0.11 °F)
≤ 0.05 °C (0.09 °F)
≤ 0.11 °C (0.2 °F)
Technische Daten
Typical measured error (±)
1)
Digital value
0.08 °C (0.14 °F)
0.06 °C (0.11 °F)
0.07 °C (0.13 °F)
Digital value
0.22 °C (0.4 °F)
1.17 °C (2.1 °F)
2.0 °C (3.6 °F)
Measured error (±)
1)
Digital
3)
Based on measured value
ME = ± (0.06 °C (0.11 °F) +
0.005% * (MV - LRV))
ME = ± (0.05 °C (0.09 °F) +
0.012% * (MV - LRV))
ME = ± (0.03 °C (0.05 °F) +
0.012% * (MV - LRV))
ME = ± (0.02 °C (0.04 °F) +
0.012% * (MV - LRV))
ME = ± (0.05 °C (0.09 °F) +
0.006% * (MV - LRV))
ME = ± (0.1 °C (0.18 °F) +
0.008% * (MV - LRV))
ME = ± (0.05 °C (0.09 °F) +
0.006% * (MV - LRV))
ME = ± (0.05 °C (0.09 °F) -
0.006% * (MV - LRV))
ME = ± (0.10 °C (0.18 °F) +
0.006% * (MV - LRV))
ME = ± (0.05 °C (0.09 °F) +
0.003% * (MV - LRV))
ME = ± (0.06 °C (0.11 °F) -
0.005% * (MV - LRV))
ME = ± (0.05 °C (0.09 °F) -
0.005% * (MV - LRV))
ME = ± (0.1 °C (0.18 °F) +
0.004% * (MV - LRV))
Value at current
output
0.1 °C (0.18 °F)
0.1 °C (0.18 °F)
0.09 °C (0.16 °F)
Value at current
output
0.24 °C (0.43 °F)
1.33 °C (2.4 °F)
2.4 °C (4.32 °F)
2)
D/A
4)
0.03 % (
4.8 µA)
51