Option Enable Utility

Enabling or Removing Options

Enabling or Removing Options
There are two types of options:
• Hardware: Hardware options involve adding additional hardware to the analyzer.
After the proper hardware has been installed in the analyzer, the option
can be enabled using the option enable utility.
It is necessary to re-enable all hardware options if the A16 test set
motherboard is replaced
• Software:
Software options add features or functionality to the analyzer without the
need for additional hardware. They are enabled using the option enable
utility.
It is necessary to re-enable all software options if the A10 frequency
reference board is replaced.
To enable or remove an option:
1. Start the option enable utility. Refer to
B-3.
2. Click the arrow in the
list below, will appear.
010 - Time Domain
UNL - Bias Tees w/Atten
014 - Configurable Test Set
016 - Receiver Attenuators
080 - Frequency Offset Mode
081 - Reference Transfer Switch
082 - Active Device Meas Appl
083 - Mixer Measurement Appl
??? - Enter Unlisted Option
3. Click on the option that you wish to either enable or remove, and then click
, whichever is appropriate.
Remove
4. If the desired option is not available in the list, select
appears that will allow you to enter the option number. Enter the option number and
follow the instructions on the display.
B-4
"Accessing the Option Enable Utility" on page
Select Desired Option
PNA Series Microwave Network Analyzers
box. A list of available options, similar to the
Enter Unlisted Option
Service Guide E8364-90026
E8362B, E8363B, E8364B
or
Enable
. A dialog box