5. If injectors are used, press the "Injector" icons to activate injectors and set
the Volume, Delay and Integration time.
6. Touch the "Door" button to automatically open the instrument door.
7. Place the sample plate on the Microplate Sample Tray. The A1 well must be
at the top right corner.
8. Touch the "Door" button again to automatically close the Microplate
Sample Tray Cover and retract the sample plate back into the instrument.
Do not push down on the Microplate Sample Plate Cover or force it to
close.
9. Touch the "Start" button.
10. To name the results file, touch the "Edit" button to activate a keypad. Enter
the results file name and touch the "OK" button when editing is complete.
11. Touch the "OK" button on the dialog screen to save. The GloMax
Detection System will initialize and begin the run.
12. To stop a run at any time, touch the "Stop" button.
13. Remove the sample plate after the run is complete.
Note: To prevent dehydration and spills, remove the sample plate from the
!
instrument when a reading is completed.

10.E. Kinetic Mode

In Kinetic mode, the instrument performs multiple reads of a single well with
defined frequency over a defined period of time before moving to the next
well. The kinetic feature is available in Luminescence mode only. One or both
injectors can be used in the Kinetic mode. Each sample reading is taken after
the second injection and injection delay.
Note: Kinetic Mode is not compatible with the 384-well plate format or Well
Scan Mode for 6- to 48-well plate formats.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Printed in USA.
Revised 5/10
®
-Multi+
Part# TM311
Page 33

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8031E8041E8061E9061E9031

Table of Contents