B. Fluorescence Module - Promega E8051 Technical Manual

Glomax multi+ detection system
Table of Contents

Advertisement

17.B. Fluorescence Module

The Fluorescence Module is required for fluorescence detection. If your
instrument does not have this module, contact Promega Technical Services for
information about purchasing and installation of the Fluorescence Module. See
Section 22.D for ordering information.
Installing the Fluorescence Module
1. If the door is open, touch the "Door" button to automatically close the
instrument door and set the Microplate Sample Tray into Home position.
2. Turn off the GloMax
3. Manually open the instrument door, and hold it open.
4. Slide the Fluorescence Detection Head into the position labeled Fluor
located inside the instrument. Be careful not to damage the printed circuit
board on the back of the detection head.
5. Hand-tighten the two captive screws using the Allen Wrench (size 7/16
inches) provided with the kit. Using too much force may damage the
screws. See Figure 44 to locate the screws, A and B.
A
B
Figure 44. Arrows A and B point to the two captive screws on the fluorescence
module.
6. Insert one of the Fluorescence Optical Kits into the opening found on the
detection head. Push in completely.
7. Close the instrument door.
8. Power on the instrument.
9. To ensure the instrument detects the newly installed module, verify the
"Fluorescence" icon appears on the Instrument Status bar on the
Instrument Control screen. See Figure 45.
Note: To prevent dust from accumulating on the optics, always keep an
Optical Kit in the Fluorescence Module and keep the instrument door
closed.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Printed in USA.
Revised 5/10
®
-Multi+ Detection System.
Part# TM311
Page 57

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8031E8041E8061E9061E9031

Table of Contents