17.D. UV-Vis Absorbance Module

To add UV-Vis detection capability to the GloMax
contact Promega Customer Services or a Distributor for information on the
purchase and installation of the UV-Vis Detection Module. See Section 22.D,
Appendix, for ordering information.
Installing the UV-Vis Absorbance Housing Unit
1. Touch the "Door" button to automatically close the instrument door and
reset the Microplate Sample Tray to Home position.
2. Power off the GloMax
supply from the back of the instrument.
3. Gently tip the GloMax
panel so that the bottom of the instrument is accessible.
Figure 51. Removing the metal cover plate.
4. Using a 3/16 inch Allen wrench remove the four 3/16 inch Allen head
screws from the metal plate cover exposing the interior compartment
(Figure 51).
5. Carefully insert the connector on UV-Vis housing unit into the connector
inside the bottom of the GloMax
Figure 52. Attaching the UV-Vis housing unit.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Printed in USA.
Revised 5/10
®
-Multi+ Detection System and disconnect the power
®
-Multi+ Detection System so that it rests on its back
®
-Multi+ Detection System.
®
-Multi+ Detection System,
Part# TM311
Page 61