20.K. Changing the Standard Light Plate Battery

1. Remove the two Phillips head screws.
2. Remove the battery retainer plate.
3. Pull the battery out of the battery compartment.
4. Install the new battery (4LR61) with notch aligned according to the
drawing.
5. Reinstall the battery retainer plate.
6. Replace the two Phillips head screws.
7. Press the "Start" button while observing the "Battery Check" indicator to
ensure the light plate is working (green light should appear).
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Printed in USA.
Revised 5/10
Part# TM311