E. Well Scan Mode - Promega E8051 Technical Manual

Glomax multi+ detection system
Table of Contents

Advertisement

Figure 25. Insert the injector tips into the injector tip holder. Snap into place.

9.E. Well Scan Mode

When using formats other than 96- and 384-well plates, the GloMax
Detection System has the option to scan each well multiple times and display
the results by showing an average of those readings.
For 6-well plates the instrument will take a total of nine readings. For 12-, 24-
and 48-well plates the instrument will take five readings (Figure 26).
Viewing of all readings per well can be accessed by zooming in on the
particular well and touching the well on the User Interface screen once the
plate read is finished (Section 9.F).
Note: Scan mode for 6- to 48-well plate formats is not compatible with Kinetics
measurements and dual-wavelength Absorbance measurements.
6-well
Figure 26. Read Patterns for Low-Density Plate Formats. The read pattern for
6-well plates (left) and for 12-, 24- and 48-well plates (right).
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Printed in USA.
Revised 5/10
12-, 24, and 48-well
®
-Multi+
Part# TM311
Page 29

Advertisement

Table of Contents
loading

This manual is also suitable for:

E8031E8041E8061E9061E9031

Table of Contents