14. Repeat Runs

Each detection modality has the option to do repeat runs. Injectors cannot be
used in this mode. In repeat runs, the entire sample plate is read between each
period.
Figure 39. Parameters for repeat runs.
14.A. Repeat Runs Parameters
Readings: The number of times a sample plate is read. When the number of
readings is set to a value between 2 and 99, the period parameter and time
estimates are displayed. The time estimate for a plate (located below the
readings parameter) is based on the number of wells selected to read.
Period: The time difference between readings of the same well. If this value is
less than the time estimated to read the plate, a caution message will appear to
inform user that the actual interval between readings will be longer. The
instrument will begin reading A1 (or the first well selected) immediately after
completing the last reading of the plate. The actual period time is indicated on
the results .csv file when exported to a computer. The range is between 1 to
120 minutes in 1-minute increments.
14.B. Viewing Repeat Runs Data on the Results Screen
The results are displayed as a set of data on the Results screen. To view the
other sets of data, touch the "Plus" button to move forward and touch the
"Minus" button to move back. All data from repeat runs are saved in one
results file. The GloMax
its internal memory. However, when the "Files" button is touched, only 50 of
the most recent files are visibly listed. To view results files not displayed,
touch the "USB" button to transfer all files to a USB flash drive and view them
on a computer.
Promega Corporation · 2800 Woods Hollow Road · Madison, WI 53711-5399 USA
Toll Free in USA 800-356-9526 · Phone 608-274-4330 · Fax 608-277-2516 · www.promega.com
Part# TM311
Page 46
®
-Multi+ Detection System saves all captured data to
Printed in USA.
Revised 5/10