9.D. Changing the Optical Crosstalk Mask for 384-Well Plates

Perform these steps only when using the 384-well plate format.
1. Remove the injector tips from the injector tip holder.
2. Remove the injector tip holder from the optical head. Facing the front of
the GloMax
®
-Multi+ Detection System, push the optical head upward, then
leftward (Figure 21).
Figure 21. Changing the optical crosstalk mask.
3. Place your hand underneath the optical mask. Grasp the sides of the optical
crosstalk mask and pull it towards you (Figure 22).
Figure 22. Grasp the sides of the optical mask and pull it towards you, being
careful to not touch the bottom of the mask.
Note: Do not touch the bottom of the optical head.